CellTrans 2018

Request Order

Fill this form to buy product. Our team will be in touch with you in less than 24h.

Description

REF 16666 (without heating block)

Transport incubator for a secure and mobile preservation of specimen, culture media etc.

Features

  • Secure transport of cells without electric mains and CO2 connection
  • Factory-aligned parameters: 37 °C and 5 % CO2
  • Different temperature or CO2 concentration possible (to be advised before order)
  • Operation at 12 V car voltage possible
  • Use of different heating blocks
  • Easy handling and cleaning
  • Compact Design


Parameters* during transport without connection to electric mains and CO
2 bottle:

Time Temp. CO2
0 min 37 °C 5 %
10 min 36.6 °C

4.9 %

30 min 35.5 °C 4.8 %
60 min 34.0 °C 4.7 %

*The measurements relate to an ambient temperature of 23 °C .

Technical Data

Power supply 115 V or 240 V AC, 50/60 Hz
12 V DC with car voltage via adapter cable
Dimensions Outside: 385 x 245 x 305 mm (l x w x h)
Inside:    approx. 312 x 167 x 140 mm (l x w x h)
Weight Approx. 4.7 kg
Temperature Available between 32 °C and 39 °C (to be advised before order)
Accuracy of regulation ± 0.3 °C (measured with heating block REF 10264)
Heating up time Approx. 2 h
CO2 concentration 5 % (different concentrations possible)
Heating block inside necessary
Connection to CO2 supply with pre-pressure of 0.8 bar necessary
CO2 accuracy ± 0.3 % CO2 (when supplied with CO2 with pre-pressure of 0.8 bar ± 0.05 bar)

Basic Supply

Basic unit
Mains cable
Car connection cable

Reviews